iXn Lux Optical Interface
![]()
The iXn iCon module drives integrated LEDs and records fluorescent responses for fiber
photometry and optogenetics applications. iX4, iX5, iX6, and iX7 devices are supported.
- iX4, iX5, and iX7 have built in light manifolds, LEDs, photosensors, and connection to an LxPM2 power meter
- iX5 and iX7 have a dedicated stim port
- iX6 drives up to three integrated LED outputs of selectable wavelengths from TDT's Lux product line and has a PM1 power meter connection
- iX4 and iX5 support two color photometry with one photosensor for fluorescent response
- iX6 and iX7 support three color photomoetry with two photosensors for fluorescent response
See the Hardware Manual for information on iX6 technical specifications.
iXn Options
| iX6 Interface |
Start by clicking the cycle button to automatically detect the Lux Pods.
| iX6 After Detect |
Then attach a Fiber Photometry gizmo to the RZ in your Processing Tree. The iXn configuration will propagate to the Fiber Photometry gizmo that it attaches to.
| iX6 Attached to Fiber Photometry Gizmo |
See the Fiber Photometry User Guide for more usage information.
iX5 and iX7 Optical Stim
The iX5 and iX7 have built-in pulse generators for driving the optical stim LED. The controls are configured through the Synapse interface. The actual pulses are generated on the iXn device itself.
| iX7 Interface with Optical Stim |