Electrical Stimulation publications

Electrical stimulation delivered with TDT stimulators: microstimulation, deep brain and nerve stimulation.

Brain Stimulation · Nov 2025

Hierarchical Bayesian estimation of motor-evoked potential recruitment curves yields accurate and robust estimates (opens in a new tab)

Vishweshwar Tyagi, Lynda M. Murray, Ahmet S. Asan, … James R. McIntosh

Abstract

Purpose: We aim to develop a robust method to improve the estimation accuracy of motor-evoked potential (MEP) recruitment curves (RCs), including motor threshold, in small-sample settings which typically involve fewer than 40 stimuli. Methods: We present a hierarchical Bayesian (HB) method to model MEP size as a rectified-logistic function of stimulation intensity. This method is designed to account for small samples, handle outliers without discarding data, quantify estimation uncertainty, and simulate synthetic data that closely matches real observations, useful for optimizing experimental design. We validate its performance on transcranial magnetic stimulation (TMS), epidural spinal cord stimulation (SCS), and synthetic TMS datasets, and provide an open-source library for Python, called hbMEP, for diverse applications. Results: The rectified-logistic outperformed sigmoidal functions in predictive accuracy on TMS and SCS datasets, as demonstrated through cross-validation. A mixture extension of the HB model improved robustness to outliers by further increasing its predictive accuracy. The HB model reduced threshold estimation error by up to 70% on sparse synthetic TMS data compared to non-hierarchical models. Bayesian estimation with the HB model reduced the required number of participants by at least 23% to detect a shift in threshold with 80% power, compared to frequentist testing. Empirical results on human SCS data further validated its applicability to real data. Conclusion: By improving accuracy on sparse data, our method minimizes the number of stimuli needed to probe each individual's neuromuscular parameters across multiple muscles simultaneously, thereby reducing session duration and the risk of inadvertent neuromodulation. Our approach provides a more statistically powerful and conclusive framework for inferring changes in threshold, and therefore corticospinal excitability. The hbMEP library streamlines and unifies the analysis of RCs across stimulation modalities and experimental paradigms.

iScience · Oct 2025

Combining cortical and spinal stimulation maximizes the improvement of gait after spinal cord injury (opens in a new tab)

Roxanne Drainville, Marco Bonizzato, Davide Burchielli, … Marina Martinez

Abstract

Most spinal cord injuries (SCI) spare descending and sublesional circuits that can be targeted through motor cortex and spinal cord stimulation to mitigate locomotor deficits. However, the potential synergy between cortical and spinal stimulation remains unknown. Here, we first investigated the phase-locked electrical stimulation of the motor cortex and lumbar spinal cord at 40 Hz in a rat model of unilateral SCI. Combining cortical and lumbar stimulation around the anticipated lift synergistically enhanced leg movements. When integrated into rehabilitation training, cortical stimulation proved essential for the recovery of skilled locomotion. We next investigated the effects of high-frequency (330 Hz) lumbar and sacral stimulation combined with cortical stimulation. Timely integration during the swing phase showed that cortical and lumbar stimulations enhance the initial and mid-swing phases, while sacral stimulation improves extension velocity in the late swing. These findings indicate that supraspinal and sublesional neuromodulation offer complementary neuroprosthetic effects in targeted SCI gait rehabilitation.

Frontiers in Neurology · Mar 2025

Epineural stimulation on distal brachial plexus for functional restoration of the upper limb in a primate study (opens in a new tab)

Tianfang Yan, Benjamin C. Fortune, Lingjun Liu, … Masayuki Hirata

Abstract

Restoring upper limb function is critical in individuals with central paralysis, and hand control is a priority in patients with neurological impairments. Functional electrical stimulation with implantable electrodes targeting the peripheral nervous system has the potential to selectively recruit hand muscles and generate multiple functional hand movements. However, the implantation of electrodes in the forearm or elbow areas requires multiple incisions for surgery, and elbow joint movements cannot be performed. In this study, we designed and implanted two epineural cuffs on the median and radial nerves in the distal brachial plexus of a single Japanese macaque ( Macaca fuscata ) monkey. The cuffs were successfully placed via an axillary approach using a single incision. Electrical stimuli were applied to innervate the contraction patterns of the hand, forearm, and triceps muscles relevant to the median and radial nerves. The evoked potentials of the target muscles electrically stimulated the distal brachial plexus to reliably and selectively innervate the upper limb muscles at the functional group level. Our results demonstrated that the distal brachial plexus can be a useful stimulation site for upper limb muscle contraction and that the axillary approach enables electrode placement to peripheral nerves required for upper limb control.

Scientific Reports · Dec 2024

An ultra low frequency spike timing dependent plasticity based approach for reducing alcohol drinking (opens in a new tab)

Anders J. Asp, Suelen Lucio Boschen, Su-Youne Chang, … J. Luis Lujan

Abstract

Alcohol use disorder (AUD) is a chronic relapsing brain disorder characterized by an impaired ability to stop or control alcohol consumption despite adverse social, occupational, or health consequences. AUD affects nearly one-third of adults at some point during their lives, with an associated cost of approximately $249 billion annually in the U.S. alone. The effects of alcohol consumption are expected to increase significantly during the COVID-19 pandemic, with alcohol sales increasing by approximately 54%, potentially exacerbating health concerns and risk-taking behaviors. Unfortunately, existing pharmacological and behavioral therapies for AUD are associated with poor success rates, with approximately 40% of individuals relapsing within three years of treatment.Pre-clinical studies have shown that chronic alcohol consumption leads to significant changes in synaptic function within the dorsal medial striatum (DMS), one of the brain regions associated with AUD and responsible for mediating goal-directed behavior. Specifically, chronic alcohol consumption has been associated with hyperactivity of dopamine receptor 1 (D1) medium spiny neurons (MSN) and hypoactivity of dopamine receptor 2 (D1) MSNs within the DMS. Optogenetic, chemogenetic, and transgenic approaches have demonstrated that reducing the D1/D2 MSN signaling imbalance decreases alcohol self-administration in rodent models of AUD.Here, we present an electrical stimulation approach that uses ultra-low (≤ 1 Hz) frequency (ULF) spike-timing-dependent plasticity (STDP) in mouse models of AUD to reduce DMS D1/D2 MSN signaling imbalances by stimulating D1-MSN afferents into the GPi and ACC glutamatergic projections to the DMS in a time-locked stimulation sequence. Our data suggest that GPi/ACC ULF-STDP selectively decreases DMS D1-MSN hyperactivity leading to reduced alcohol consumption without evoking undesired affective behaviors using electrical stimulation rather than approaches requiring genetic modification. This work represents a step towards fulfilling the unmet need for a reliable method of treating severe AUD through cell-type-specific control with clinically available neuromodulation tools.

Sensors · Oct 2024

Comparison of Subdural and Intracortical Recordings of Somatosensory Evoked Responses (opens in a new tab)

Felipe Rettore Andreis, Suzan Meijs, Thomas Gomes Nørgaard dos Santos Nielsen, … Winnie Jensen

Abstract

Micro-electrocorticography (µECoG) electrodes have emerged to balance the trade-off between invasiveness and signal quality in brain recordings. However, its large-scale applicability is still hindered by a lack of comparative studies assessing the relationship between ECoG and traditional recording methods such as penetrating electrodes. This study aimed to compare somatosensory evoked potentials (SEPs) through the lenses of a µECoG and an intracortical microelectrode array (MEA). The electrodes were implanted in the pig’s primary somatosensory cortex, while SEPs were generated by applying electrical stimulation to the ulnar nerve. The SEP amplitude, signal-to-noise ratio (SNR), power spectral density (PSD), and correlation structure were analysed. Overall, SEPs resulting from MEA recordings had higher amplitudes and contained significantly more spectral power, especially at higher frequencies. However, the SNRs were similar between the interfaces. These results demonstrate the feasibility of using µECoG to decode SEPs with wide-range applications in physiology monitoring and brain–computer interfaces.

Nature Communications · Oct 2024

Border-ownership tuning determines the connectivity between V4 and V1 in the macaque visual system (opens in a new tab)

Danique Jeurissen, Anne F. van Ham, Amparo Gilhuis, … Matthew W. Self

Abstract

Cortical feedback connections are extremely numerous but the logic of connectivity between higher and lower areas remains poorly understood. Feedback from higher visual areas to primary visual cortex (V1) has been shown to enhance responses on perceptual figures compared to backgrounds, an effect known as figure-background modulation (FBM). A likely source of this feedback are border-ownership (BO) selective cells in mid-tier visual areas (e.g. V4) which represent the location of figures. We examined the connectivity between V4 cells and V1 cells using noise-correlations and micro-stimulation to estimate connectivity strength. We show that connectivity is consistent with a model in which BO-tuned V4 cells send positive feedback in the direction of their preferred figure and negative feedback in the opposite direction. This connectivity scheme can recreate patterns of FBM observed in previous studies. These results provide insights into the cortical connectivity underlying figure-background perception and establish a link between FBM and BO-tuning.

Cognitive, Affective, & Behavioral Neuroscience · Aug 2024

Cortico-striatal beta oscillations as a reward-related signal (opens in a new tab)

M. F. Koloski, S. Hulyalkar, S. A. Barnes, … D. S. Ramanathan

Abstract

The value associated with reward is sensitive to external factors, such as the time between the choice and reward delivery as classically manipulated in temporal discounting tasks. Subjective preference for two reward options is dependent on objective variables of reward magnitude and reward delay. Single neuron correlates of reward value have been observed in regions, including ventral striatum, orbital, and medial prefrontal cortex. Brain imaging studies show cortico-striatal-limbic network activity related to subjective preferences. To explore how oscillatory dynamics represent reward processing across brain regions, we measured local field potentials of rats performing a temporal discounting task. Our goal was to use a data-driven approach to identify an electrophysiological marker that correlates with reward preference. We found that reward-locked oscillations at beta frequencies signaled the magnitude of reward and decayed with longer temporal delays. Electrodes in orbitofrontal/medial prefrontal cortex, anterior insula, ventral striatum, and amygdala individually increased power and were functionally connected at beta frequencies during reward outcome. Beta power during reward outcome correlated with subjective value as defined by a computational model fit to the discounting behavior. These data suggest that cortico-striatal beta oscillations are a reward signal correlated, which may represent subjective value and hold potential to serve as a biomarker and potential therapeutic target.

STAR Protocols · Mar 2024

Gaussian-process-based Bayesian optimization for neurostimulation interventions in rats (opens in a new tab)

Léo Choinière, Rose Guay-Hottin, Rémi Picard, … Numa Dancause

Abstract

Effective neural stimulation requires adequate parametrization. Gaussian-process (GP)-based Bayesian optimization (BO) offers a framework to discover optimal stimulation parameters in real time. Here, we first provide a general protocol to deploy this framework in neurostimulation interventions and follow by exemplifying its use in detail. Specifically, we describe the steps to implant rats with multi-channel electrode arrays in the hindlimb motor cortex. We then detail how to utilize the GP-BO algorithm to maximize evoked target movements, measured as electromyographic responses. For complete details on the use and execution of this protocol, please refer to Bonizzato and colleagues (2023).1.

IEEE Transactions on Neural Systems and Rehabilitation Engineering · 2024

Fabrication and Validation of Sub-Cellular Carbon Fiber Electrodes (opens in a new tab)

Julianna Richie, Joseph G. Letner, Autumn Mclane-Svoboda, … Cynthia A. Chestek

Abstract

Multielectrode arrays for interfacing with neurons are of great interest for a wide range of medical applications. However, current electrodes cause damage over time. Ultra small carbon fibers help to address issues but controlling the electrode site geometry is difficult. Here we propose a methodology to create small, pointed fiber electrodes (SPFe). We compare the SPFe to previously made blowtorched fibers in characterization. The SPFe result in small site sizes [Formula: see text] with consistently sharp points (20.8 ± 7.64°). Additionally, these electrodes were able to record and/or stimulate neurons multiple animal models including rat cortex, mouse retina, Aplysia ganglia and octopus axial cord. In rat cortex, these electrodes recorded significantly higher peak amplitudes than the traditional blowtorched fibers. These SPFe may be applicable to a wide range of applications requiring a highly specific interface with individual neurons.

Nature Communications · Jul 2023

Conductance stable and mechanically durable bi-layer EGaIn composite-coated stretchable fiber for 1D bioelectronics (opens in a new tab)

Gun-Hee Lee, Do Hoon Lee, Woojin Jeon, … Steve Park

Abstract

Deformable semi-solid liquid metal particles (LMP) have emerged as a promising substitute for rigid conductive fillers due to their excellent electrical properties and stable conductance under strain. However, achieving a compact and robust coating of LMP on fibers remains a persistent challenge, mainly due to the incompatibility of conventional coating techniques with LMP. Additionally, the limited durability and absence of initial electrical conductivity of LMP restrict their widespread application. In this study, we propose a solution process that robustly and compactly assembles mechanically durable and initially conductive LMP on fibers. Specifically, we present a shearing-based deposition of polymer-attached LMP followed by additional coating with CNT-attached LMP to create bi-layer LMP composite with exceptional durability, electrical conductivity, stretchability, and biocompatibility on various fibers. The versatility and reliability of this manufacturing strategy for 1D electronics are demonstrated through the development of sewn electrical circuits, smart clothes, stretchable biointerfaced fiber, and multifunctional fiber probes.

Frontiers in Neuroscience · Apr 2023

Layer-dependent stability of intracortical recordings and neuronal cell loss (opens in a new tab)

Morgan E. Urdaneta, Nicolas G. Kunigk, Jesus D. Peñaloza-Aponte, … Kevin J. Otto

Abstract

Intracortical recordings can be used to voluntarily control external devices via brain-machine interfaces (BMI). Multiple factors, including the foreign body response (FBR), limit the stability of these neural signals over time. Current clinically approved devices consist of multi-electrode arrays with a single electrode site at the tip of each shank, confining the recording interface to a single layer of the cortex. Advancements in manufacturing technology have led to the development of high-density electrodes that can record from multiple layers. However, the long-term stability of neural recordings and the extent of neuronal cell loss around the electrode across different cortical depths have yet to be explored. To answer these questions, we recorded neural signals from rats chronically implanted with a silicon-substrate microelectrode array spanning the layers of the cortex. Our results show the long-term stability of intracortical recordings varies across cortical depth, with electrode sites around L4-L5 having the highest stability. Using machine learning guided segmentation, our novel histological technique, DeepHisto, revealed that the extent of neuronal cell loss varies across cortical layers, with L2/3 and L4 electrodes having the largest area of neuronal cell loss. These findings suggest that interfacing depth plays a major role in the FBR and long-term performance of intracortical neuroprostheses.

IBRO Neuroscience Reports · Dec 2021

The effect of peripheral high-frequency electrical stimulation on the primary somatosensory cortex in pigs (opens in a new tab)

Taha Al Muhammadee Janjua, Thomas Gomes Nørgaard dos Santos Nielsen, Felipe Rettore Andreis, … Winnie Jensen

Abstract

This study implements the use of Danish Landrace pigs as subjects for the long-term potentiation (LTP)-like pain model. This is accomplished by analyzing changes in the primary somatosensory cortex (S1) in response to electrical stimulation on the ulnar nerve after applying high-frequency electrical stimulation (HFS) on the ulnar nerve. In this study, eight Danish Landrace pigs were electrically stimulated, through the ulnar nerve, to record the cortically evoked response in S1 by a 16-channel microelectrode array (MEA). Six of these pigs were subjected to HFS (four consecutive, 15 mA, 100 Hz, 1000 µs pulse duration) 45 min after the start of the experiment. Two pigs were used as control subjects to compare the cortical response to peripheral electrical stimulation without applying HFS. Low-frequency components of the intracortical signals (0.3-300 Hz) were analyzed using event-related potential (ERP) analysis, where the minimum peak during the first 30-50 ms (N1 component) in each channel was detected. The change in N1 was compared over time across the intervention and control groups. Spectral analysis was used to demonstrate the effect of the intervention on the evoked cortical oscillations computed between 75 ms and 200 ms after stimulus. ERP analysis showed an immediate increase in N1 amplitude that became statistically significant 45 mins after HFS (p < 0.01) for the intervention group. The normalized change in power in frequency oscillations showed a similar trend. The results show that the LTP-like pain model can be effectively implemented in pigs using HFS since the cortical responses are comparable to those described in humans.

Journal of NeuroEngineering and Rehabilitation · May 2021

Epidural cerebellar stimulation drives widespread neural synchrony in the intact and stroke perilesional cortex (opens in a new tab)

Aamir Abbasi, Nathan P. Danielsen, Jennifer Leung, … Tanuj Gulati

Abstract

Background Cerebellar electrical stimulation has shown promise in improving motor recovery post-stroke in both rodent and human studies. Past studies have used motor evoked potentials (MEPs) to evaluate how cerebellar stimulation modulates ongoing activity in the cortex, but the underlying mechanisms are incompletely understood. Here we used invasive electrophysiological recordings from the intact and stroke-injured rodent primary motor cortex (M1) to assess how epidural cerebellar stimulation modulates neural dynamics at the level of single neurons as well as at the level of mesoscale dynamics. Methods We recorded single unit spiking and local field potentials (LFPs) in both the intact and acutely stroke-injured M1 contralateral to the stimulated cerebellum in adult Long-Evans rats under anesthesia. We analyzed changes in the firing rates of single units, the extent of synchronous spiking and power spectral density (PSD) changes in LFPs during and post-stimulation. Results Our results show that post-stimulation, the firing rates of a majority of M1 neurons changed significantly with respect to their baseline rates. These firing rate changes were diverse in character, as the firing rate of some neurons increased while others decreased. Additionally, these changes started to set in during stimulation. Furthermore, cross-correlation analysis showed a significant increase in coincident firing amongst neuronal pairs. Interestingly, this increase in synchrony was unrelated to the direction of firing rate change. We also found that neuronal ensembles derived through principal component analysis were more active post-stimulation. Lastly, these changes occurred without a significant change in the overall spectral power of LFPs post-stimulation. Conclusions Our results show that cerebellar stimulation caused significant, long-lasting changes in the activity patterns of M1 neurons by altering firing rates, boosting neural synchrony and increasing neuronal assemblies’ activation strength. Our study provides evidence that cerebellar stimulation can directly modulate cortical dynamics. Since these results are present in the perilesional cortex, our data might also help explain the facilitatory effects of cerebellar stimulation post-stroke.

Frontiers in Neuroscience · Mar 2021

Improving the Selectivity of an Osseointegrated Neural Interface: Proof of Concept For Housing Sieve Electrode Arrays in the Medullary Canal of Long Bones (opens in a new tab)

Augusto X. T. Millevolte, Aaron M. Dingle, Jared P. Ness, … Samuel O. Poore

Abstract

Sieve electrodes stand poised to deliver the selectivity required for driving advanced prosthetics but are considered inherently invasive and lack the stability required for a chronic solution. This proof of concept experiment investigates the potential for the housing and engagement of a sieve electrode within the medullary canal as part of an osseointegrated neural interface (ONI) for greater selectivity toward improving prosthetic control. The working hypotheses are that (A) the addition of a sieve interface to a cuff electrode housed within the medullary canal of the femur as part of an ONI would be capable of measuring efferent and afferent compound nerve action potentials (CNAPs) through a greater number of channels; (B) that signaling improves over time; and (C) that stimulation at this interface generates measurable cortical somatosensory evoked potentials through a greater number of channels. The modified ONI was tested in a rabbit ( n = 1) amputation model over 12 weeks, comparing the sieve component to the cuff, and subsequently compared to historical data. Efferent CNAPs were successfully recorded from the sieve demonstrating physiological improvements in CNAPs between weeks 3 and 5, and somatosensory cortical responses recorded at 12 weeks postoperatively. This demonstrates that sieve electrodes can be housed and function within the medullary canal, demonstrated by improved nerve engagement and distinct cortical sensory feedback. This data presents the conceptual framework for housing more sophisticated sieve electrodes in bone as part of an ONI for improving selectivity with percutaneous connectivity toward improved prosthetic control.

Journal of NeuroEngineering and Rehabilitation · Jan 2021

Discriminability of multiple cutaneous and proprioceptive hand percepts evoked by intraneural stimulation with Utah slanted electrode arrays in human amputees (opens in a new tab)

David M. Page, Jacob A. George, Suzanne M. Wendelken, … Gregory A. Clark

Abstract

Background Electrical stimulation of residual afferent nerve fibers can evoke sensations from a missing limb after amputation, and bionic arms endowed with artificial sensory feedback have been shown to confer functional and psychological benefits. Here we explore the extent to which artificial sensations can be discriminated based on location, quality, and intensity. Methods We implanted Utah Slanted Electrode Arrays (USEAs) in the arm nerves of three transradial amputees and delivered electrical stimulation via different electrodes and frequencies to produce sensations on the missing hand with various locations, qualities, and intensities. Participants performed blind discrimination trials to discriminate among these artificial sensations. Results Participants successfully discriminated cutaneous and proprioceptive sensations ranging in location, quality and intensity. Performance was significantly greater than chance for all discrimination tasks, including discrimination among up to ten different cutaneous location-intensity combinations (15/30 successes, p < 0.0001) and seven different proprioceptive location-intensity combinations (21/40 successes, p < 0.0001). Variations in the site of stimulation within the nerve, via electrode selection, enabled discrimination among up to five locations and qualities (35/35 successes, p < 0.0001). Variations in the stimulation frequency enabled discrimination among four different intensities at the same location (13/20 successes, p < 0.0005). One participant also discriminated among individual stimulation of two different USEA electrodes, simultaneous stimulation on both electrodes, and interleaved stimulation on both electrodes (20/24 successes, p < 0.0001). Conclusion Electrode location, stimulation frequency, and stimulation pattern can be modulated to evoke functionally discriminable sensations with a range of locations, qualities, and intensities. This rich source of artificial sensory feedback may enhance functional performance and embodiment of bionic arms endowed with a sense of touch.

Materials · Dec 2018

Glassy Carbon Electrocorticography Electrodes on Ultra-Thin and Finger-Like Polyimide Substrate: Performance Evaluation Based on Different Electrode Diameters (opens in a new tab)

Maria Vomero, Elena Zucchini, Emanuela Delfino, … Thomas Stieglitz

Abstract

Glassy carbon (GC) has high potential to serve as a biomaterial in neural applications because it is biocompatible, electrochemically inert and can be incorporated in polyimide-based implantable devices. Miniaturization and applicability of GC is, however, thought to be partially limited by its electrical conductivity. For this study, ultra-conformable polyimide-based electrocorticography (ECoG) devices with different-diameter GC electrodes were fabricated and tested in vitro and in rat models. For achieving conformability to the rat brain, polyimide was patterned in a finger-like shape and its thickness was set to 8 µm. To investigate different electrode sizes, each ECoG device was assigned electrodes with diameters of 50, 100, 200 and 300 µm. They were electrochemically characterized and subjected to 10 million biphasic pulses—for achieving a steady-state—and to X-ray photoelectron spectroscopy, for examining their elemental composition. The electrodes were then implanted epidurally to evaluate the ability of each diameter to detect neural activity. Results show that their performance at low frequencies (up to 300 Hz) depends on the distance from the signal source rather than on the electrode diameter, while at high frequencies (above 200 Hz) small electrodes have higher background noises than large ones, unless they are coated with poly(3,4-ethylenedioxythiophene)-poly(styrenesulfonate) (PEDOT:PSS).

bioRxiv · May 2018

Optic nerve intraneural stimulation allows selective visual cortex activation (opens in a new tab)

Vivien Gaillet, Annarita Cutrone, Paola Vagni, … Diego Ghezzi

Abstract

Retinal prostheses have been developed to restore a functional form of vision in patients affected by outer retinal layer dystrophies. Other approaches, namely optic nerve, thalamic, and cortical prostheses are under investigation to expand this toolbox both as clinical devices and as tools for fundamental research. Optic nerve stimulation is attractive since it bypasses the entire retinal network and directly activates nerve fibers. At the same time, it still takes advantage of the high-level information processing occurring downstream in the visual cortex. Here we show that a new intraneural electrode array (OpticSELINE) is effective in inducing the activation of the visual cortex upon electrical stimulation of the optic nerve. We also demonstrate that intraneural optic nerve stimulation induces selective cortical activation patterns depending on the stimulating electrode, thus suggesting that the OpticSELINE possesses spatial selectivity in fiber stimulation. In conclusion, the OpticSELINE can be used both as visual prosthesis for blind patients and as tool to further investigate the effect of the electrical stimulation in the visual system.

Journal of NeuroEngineering and Rehabilitation · Nov 2017

Restoration of motor control and proprioceptive and cutaneous sensation in humans with prior upper-limb amputation via multiple Utah Slanted Electrode Arrays (USEAs) implanted in residual peripheral arm nerves (opens in a new tab)

Suzanne Wendelken, David M. Page, Tyler Davis, … Gregory A. Clark

Abstract

Background: Despite advances in sophisticated robotic hands, intuitive control of and sensory feedback from these prostheses has been limited to only 3-degrees-of-freedom (DOF) with 2 sensory percepts in closed-loop control. A Utah Slanted Electrode Array (USEA) has been used in the past to provide up to 81 sensory percepts for human amputees. Here, we report on the advanced capabilities of multiple USEAs implanted in the residual peripheral arm nerves of human amputees for restoring control of 5 DOF and sensation of up to 131 proprioceptive and cutaneous hand sensory percepts. We also demonstrate that USEA-restored sensory percepts provide a useful source of feedback during closed-loop virtual prosthetic hand control. Methods: Two 100-channel USEAs were implanted for 4-5 weeks, one each in the median and ulnar arm nerves of two human subjects with prior long-duration upper-arm amputations. Intended finger and wrist positions were decoded from neuronal firing patterns via a modified Kalman filter, allowing subjects to control many movements of a virtual prosthetic hand. Additionally, USEA microstimulation was used to evoke numerous sensory percepts spanning the phantom hand. Closed-loop control was achieved by stimulating via an electrode of the ulnar-nerve USEA while recording and decoding movement via the median-nerve USEA. Results: Subjects controlled up to 12 degrees-of-freedom during informal, 'freeform' online movement decode sessions, and experienced up to 131 USEA-evoked proprioceptive and cutaneous sensations spanning the phantom hand. Independent control was achieved for a 5-DOF real-time decode that included flexion/extension of the thumb, index, middle, and ring fingers, and the wrist. Proportional control was achieved for a 4-DOF real-time decode. One subject used a USEA-evoked hand sensation as feedback to complete a 1-DOF closed-loop virtual-hand movement task. There were no observed long-term functional deficits due to the USEA implants. Conclusions: Implantation of high-channel-count USEAs enables multi-degree-of-freedom control of virtual prosthetic hand movement and restoration of a rich selection of both proprioceptive and cutaneous sensory percepts spanning the hand during the short 4-5 week post-implant period. Future USEA use in longer-term implants and in closed-loop may enable restoration of many of the capabilities of an intact hand while contributing to a meaningful embodiment of the prosthesis.

Scientific Reports · Jan 2017

Highly Stable Glassy Carbon Interfaces for Long-Term Neural Stimulation and Low-Noise Recording of Brain Activity (opens in a new tab)

Maria Vomero, Elisa Castagnola, Francesca Ciarpella, … Davide Ricci

Abstract

We report on the superior electrochemical properties, in-vivo performance and long term stability under electrical stimulation of a new electrode material fabricated from lithographically patterned glassy carbon. For a direct comparison with conventional metal electrodes, similar ultra-flexible, micro-electrocorticography (μ-ECoG) arrays with platinum (Pt) or glassy carbon (GC) electrodes were manufactured. The GC microelectrodes have more than 70% wider electrochemical window and 70% higher CTC (charge transfer capacity) than Pt microelectrodes of similar geometry. Moreover, we demonstrate that the GC microelectrodes can withstand at least 5 million pulses at 0.45 mC/cm 2 charge density with less than 7.5% impedance change, while the Pt microelectrodes delaminated after 1 million pulses. Additionally, poly(3,4-ethylenedioxythiophene)-poly(styrenesulfonate) (PEDOT-PSS) was selectively electrodeposited on both sets of devices to specifically reduce their impedances for smaller diameters (<60 μm). We observed that PEDOT-PSS adhered significantly better to GC than Pt, and allowed drastic reduction of electrode size while maintaining same amount of delivered current. The electrode arrays biocompatibility was demonstrated through in-vitro cell viability experiments, while acute in vivo characterization was performed in rats and showed that GC microelectrode arrays recorded somatosensory evoked potentials (SEP) with an almost twice SNR (signal-to-noise ratio) when compared to the Pt ones.

Scientific Reports · Sep 2016

Flexible Neural Electrode Array Based-on Porous Graphene for Cortical Microstimulation and Sensing (opens in a new tab)

Yichen Lu, Hongming Lyu, Andrew G. Richardson, … Duygu Kuzum

Abstract

Neural sensing and stimulation have been the backbone of neuroscience research, brain-machine interfaces and clinical neuromodulation therapies for decades. To-date, most of the neural stimulation systems have relied on sharp metal microelectrodes with poor electrochemical properties that induce extensive damage to the tissue and significantly degrade the long-term stability of implantable systems. Here, we demonstrate a flexible cortical microelectrode array based on porous graphene, which is capable of efficient electrophysiological sensing and stimulation from the brain surface, without penetrating into the tissue. Porous graphene electrodes show superior impedance and charge injection characteristics making them ideal for high efficiency cortical sensing and stimulation. They exhibit no physical delamination or degradation even after 1 million biphasic stimulation cycles, confirming high endurance. In in vivo experiments with rodents, same array is used to sense brain activity patterns with high spatio-temporal resolution and to control leg muscles with high-precision electrical stimulation from the cortical surface. Flexible porous graphene array offers a minimally invasive but high efficiency neuromodulation scheme with potential applications in cortical mapping, brain-computer interfaces, treatment of neurological disorders, where high resolution and simultaneous recording and stimulation of neural activity are crucial.