Processor Comparison | RX5 Pentusa Base Station | RX6 MultiFunction Processor | RX7 Mcrostimulator Base Station | RZ2 Processor

 

 

 

 

 

RX8

RX8 Multi I/O Processor

 

 

 

 

 

   
 
   

The Multi I/O Processor (RX8) is a high channel count analog input/output system. Available with either 12 or 24 analog channels, the RX8 can be configured to your specifications at the factory. Combine PCM or Sigma-Delta A/Ds and D/As to create a system matched to your application. The Sigma-Delta converters have a dynamic range of over 100 dB with a maximum bandwidth of 50 kHz, making them ideal for audio applications. The PCM converters, with almost no group delay, are excellent for controlling external devices such as microstimulators or motors in real-time. Combine the RX8 with the RX5 for a complete physiology workstation or consider the RX6 for auditory applications that require higher sampling rates. A complete line of transducers and amplifiers are available to round out the system.

 

Part numbers:
RX8-2 Two DSPs and 12 I/O
RX8-2-24 Two DSPs and 24 I/O
RX8-5 Five DSPs and 24 I/O
RX8-5-12 Five DSPs and 12 I/O

 
   

 

Technical specifications:

DSPs

two or five 100 MHz Sharc ADSP 21161

Memory: 128 MB SDRAM
Max sampling rate: ~100 kHz
D/A up to 24 channels
Frequency response: DC-Nyquist (~1/2 sample rate)
PCM: 16-bit  

S/N (typical):

80 dB (20 Hz to 20 kHz at 9.9 V)

THD (typical):

- 70 dB for 1 kHz output at 5 Vrms

Sample delay:

4 samples
Sigma-Delta: 24-bit  

S/N (typical):

97 dB (20 Hz - 20 kHz at 9.9 V)

THD (typical):

- 84 dB for 1 kHz output at 5 Vrms

Sample delay:

23 samples
A/D up to 16 channels
Frequency response: DC-Nyquist (~1/2 sample rate)
PCM: 16-bit  

S/N (typical):

80 dB (20 Hz - 20 kHz at 9.9 V)

Sample delay:

4 samples
Sigma-Delta: 24-bit  

S/N (typical):

97 dB (20 Hz - 20 kHz at 9.9 V)

THD (typical):

- 84 dB for 1 kHz input at 5 Vrms

Sample delay:

47 samples
Digital I/O: 24 bits programmable
Power and communication: zBus required